RN172plus Gas Data Logger


Hubungi Kami





Description

The RN172plus device is designed to transmit the measured gases or thermal data via WiFI to the destination. Thedestination would be an information system such as public cloud, local server, and even PC. The RN172plus device has a display on the front to show the data value and it has an alarm output with the range threshold values. Data can be transmitted to the user's server through MODBUS TCP and HTTP communication. In addition, with the cloud-based web monitoring service Radionode365, you can use a real-time dashboard, emergency SMS/Voice alarm, report generation and mail delivery scheduling function.


Feature

  • Support Variable Sensor cartridges
  • Support WiFi (WiFi IEEE 802.11 b,g)
  • 4 Digit LED Display, 2 Lines
  • Support MODBUS TCP, HTTP, Cloud Storage
  • Alerts (Text Message, Voice Call)
  • Internal Sound Buzzer

Spesification

Measured substance

Cartridge Model

Measurement Range

VOC

UA50

CH1: 0 ~ 60,000 ppb

CH2: Absolute humidity(0 ~ 256 g/m3) or CO2eq(400 ~ 10000 ppm)

Carbon dioxide

UA52-CO2

CH1: 0~20%

CH2: Temp. 0~50?

Oxygen

UA52-O2-A

CH1: 0~25%

CH2: Temp(-30~60?) or Barometric pressure(500 ~ 1200 mBar)

Carbon monoxide

UA53-CO

CH1: 0~1000 ppm

CH2: Temp. -20~40?

Nitrogen dioxide

UA53-NO2

CH1: 0~10 ppm

CH2: Temp. -20~40?

Sulfur dioxide

UA53-SO2

CH1: 0~20 ppm

CH2: Temp. -20~40?

Ethylene

UA54-C2H4-10

CH1: 0~10ppm

CH2: Temp. -40~50?

Ethylene

UA54-C2H4-100

CH1: 0~100ppm

CH2: Temp. -40~50?

Ethylene oxide

UA54-EO

CH1: 0~10ppm

CH2: Temp. -40~50?

Hydrogen sulfide

UA54-H2S

CH1: 0~50ppm

CH2: Temp

Hydrogen chloride

UA54-HCL

CH1: 0~20ppm

CH2: Temp

Ammonia

UA54-NH3-100

CH1: 0~100ppm

CH2: Temp. -10~50?

Ammonia

UA54-NH3-1000

CH1: 0~1,000ppm

CH2: Temp. -10~50?

Oxygen

UA54-O2

CH1: 0.1~20.9%

CH2: Temp. -30~55?

Hydrogen

UA54-H2-4

CH1: 0~40,000ppm

CH2:Temp. -40~50?